Semi-empirical Calibration Method

For routine environmental sample analyses, efficiency calibration of gamma detec-

tors can be performed using (a) multiple standard point sources or standard

volume sources of mixed radionuclides (for soils measurements), or (b) filters

spiked with standard solution (for

7

Be precipitate analysis). Considering the



short half-life of

7

Be, standard sources containing



7

Be are not commercially

available. When purchasing a standard for efficiency calibration, it is essential

that these should carry a certificate of calibration from a metrological labora-




34