5.2 UV Light Source
The UV lamp is located inside the operation area to ensure that the ultraviolet light can fully 
illuminate every place in the operation area, so that the operation area is fully disinfected.
6


B IO B A 56
5.3 Lighting source
The lighting source adopts LED lights, which can ensure that the average illuminance in the 
operating area meets the standard requirements.
5.4 Waterproof socket
A waterproof socket (IP44) is installed in the operating area, which can supply power to the 
equipment in the operating area.
Note:
1) The total power of the equipment used on the two sockets does not exceed 500W (rated 
voltage*rated current 220V*2.3A);
2) The waterproof socket can only be waterproof when the front cover is down. When the front 
cover is opened, the socket cannot be regarded as a waterproof socket;
3) IP44 is the socket protection level. According to the provisions of GB4208: prevent fixed foreign 
objects with a diameter greater than 1.0mm from entering the socket shell to contact internal parts; 
prevent splashing water from all directions from entering the socket and causing damage to it.
5.5 Control Panel
Please refer to Figure 2 for the specific use of the control panel.
7


BIOBASG
a. Gear display
Through the gear display
FAN , you can understand the wind speed gear situation
when the laminar flow cabinet is working, 
b. Touch the button
The operation of the laminar flow cabinet needs to be realized by tapping the buttons. Please refer to 
Table 2 for the details of the tapping buttons.
Table 2. Detailed description of the touch button
\
Power
button
The main switch that controls other function 
button
I
©
t
1
Fan button
It is the control key for the working status of the 
fan. Each time you press it, the working status 
of the fan and the corresponding indication 
status on the LCD display will change once 
(this key is only valid when the current window 
glass is open).
H
fwlh 
]
'
Lighting
button
Each time you press, the state of the lighting 
tube changes once, that is, from off to bright, 
and from bright to off.
i
UV lamp 
button
Control button for UV lamp. Each time you 
press it, the status of the UV lamp and the 
corresponding indication status on the LCD 
display will change once, either from bright to 
off, or from off to bright (this key is valid only 
when the current window glass, is completely
i
о
Socket
button
The control key of the socket on and off the 
working state. Each time you press, the on-off 
status and the corresponding indication status 
on the LCD display will change once.
Function
setting
debugging
button
Use this button when debugging and setting 
related functions.