Fractional laser system


Download 81.25 Kb.
Pdf ko'rish
bet5/10
Sana08.09.2023
Hajmi81.25 Kb.
#1674024
1   2   3   4   5   6   7   8   9   10
Bog'liq
CO2 fractional Vaginal tightenning laser user\'s manual

7. Direction for Use
7.1Structure description
1. Joints of six light guide arm
The laser beam transmission is completed by the six-joint light guide arm.
The role of the light guide arm is to transmit the laser to the patient's site
(target surface) to be operated on
2. Probe
Graphics scanner can scan output, rectangular, round, hexagonal,
triangular and other graphics, the largest area of about 20 * 20mm² scanning;
3. Main unit
It includes a high voltage power supply, a low-voltage power supply,
control system, cooling system, laser system and blowing system.


4. touch display panel
Mainly used for parameter setting
5. Emergency switch
When the instrument has an abnormal condition, you can press the
emergency stop switch to cut off the power supply system of the instrument
6.key switch
A key into the key switch, go to the "open" position, the instrument power on,
go to the "off” position to cut the power, turn off the instrument.
7.Affusion hole
8.Overflow hole
9. Drainage hole


10.Foot switch
11.The power input socket
7.2 Installation
1) Checking
1. Check the packing box before open, if no abnormal, you can open it.
Becase the machine is equipped with laser tube and precision optical system,
forbid pound and collision when open the box.
2. Take out the packing list after opening the box, then take out the host and
spare parts
according to the packing list, check every spare parts, wire connections,
confirmed no
abnormal or troubleshooting before installation, if have any problem, please
contact
manufacturer timely, so that resolve the problem Quickly.
2) Fill distilled water
1. Press the stainless steel sheet at the above of the overflow hole, pull the
water tube into the hole.
2. Put the tube into the affusion hole, do like as above said, fill distilled water,
it's ok when
the water overflow the overflow hole; after turn on the machine for 3 minutes,
the machine need fill water again.
Attention: you can't pull out the tube from the affusion hole when using the
machine. You need to change the water in every 60 days.

Download 81.25 Kb.