5.6 Fuse
The laminar flow cabinet is equipped with a power fuse, which is located on the side of the power 
cord outlet on the side of the box. For the corresponding specifications and models of the fuse, please 
refer to the section 9. Label description.
6. Installation and Operation Instructions
6.1 Installation
a. Remove all packaging components.
b. Check whether there are scratches, deformations or foreign objects on the outer surface of the 
host.
c. Check the relevant accessories and information carefully according to the packing list of the 
accessories in the manual.
d. Move the equipment near the final placement point and facilitate the installation position, prepare 
the S8 Allen wrench that needs to be used, and then proceed with the installation operation.
Note:
The base part is placed at the back o f the laminar flow cabinet during transportation, and it 
should be taken out before installation; it is strictly prohibited to move the laminar flow 
cabinet with the bottom facing upwards, and it is strictly forbidden to place or disassemble it.
e. Assembly of base part
Before assembling the base, please unscrew the M l0*55 hexagon socket head bolts in the cross 
brace and T-bracket. For the specific assembly process of the'base, please refer to the installation 
diagram of the base, as shown in Figure 3.
Figure 3
10


B IO BAS6
f. Adjustment of Footmaster Caster
i;
Clockwise: lower the cabinet, move the 
cabinet
Counterclockwise: increase the cabinet, tix 
the cabinet
Figure 4
Please refer to Figure 4. Rotate the red part of the Fuma casters clockwise to lower the corresponding 
base feet, thereby reducing the height of the cabinet. When the four Fuma casters are lowered at the 
same time, the cabinet can be moved; Rotate the red part of the Fuma casters counterclockwise to 
raise the corresponding base feet to increase the height of the cabinet. When the four Fuma casters 
are raised at the same time, the cabinet can be fixed; The simultaneous adjustment of four Fuma 
casters can make the cabinet in a horizontal and stable state,
g. The host is connected to the base
Please refer to Figure 5 for the connection between the host and the base. To connect the host to the 
base, M l0*20 hexagon socket head bolts, flat washers 10, and spring washers 10 need to pass 
through the base from the bottom up to the host side plate to fix the host and the base firmly.
M10 Flat washers 
M10 Spring washer
(4pcs)
Use bolts to connect the base and the host
Figure 5
h. Placement of the equipment
The laminar flow cabinet should be placed in a relatively stable indoor airflow area to reduce the 
impact of environmental factors such as ventilation system, air conditioning, doors, windows, and
n


BIOBASG
personnel movement on the performance of the clean workbench; It is forbidden to place the laminar 
flow cabinet in a location that is difficult to cut off, and at least 300mm of space must be reserved on 
each side of the laminar flow cabinet for subsequent inspection and maintenance.