laboratory. The ORS laboratory protocol is summarized below for 
convenience. 
Perform medical device extractions and final endotoxin concentration 
calculations per USP <161> and USP <1085> Guidelines on the Endotoxins 
Test. 
Alternatively, if deemed necessary, perform the ORS laboratory modified 
extraction and analysis of Endotoxin from Medical Devices: 
a. Preparation of 1% SLS solution 
Prepare a 1% stock solution by placing one (1) gram of sodium lauryl sulfate 
(SLS) into a depyrogenated glass flask and add 99 ml of pyrogen-free water. 
Allow the SLS to fully dissolve. This should be followed by filtration through a 
10,000 MW depyrogenation membrane filter into a pyrogen-free glass or 
plastic container. 
b. Equipment needed: Ultrasonic bath with a range of 150 to 480 
watts.
c. Extraction procedure 
i. Dilute 2mL of 1% SLS stock solution to 20mL (0.1%) using 
LAL reagent water in a 20 x 150 mm screw-cap tube. 
ii. Dilute 1.5mL of the 0.1% SLS solution to 15 mL (0.01%) using 
LAL reagent water in a 20 x 150 mm screw-cap tube. 
iii. Prepare the appropriate number of tubes (one tube for each 
device) and one as a negative / system control. Preheat in a 
water bath to 37°C ±1°C. 
iv. Aseptically remove the device from its packaging and cut it 
diagonally into pieces less than 5mm in length. Small metal 
and plastic pieces such as needles and luer-locks should be 
tested whole. Pyrogen-free fluid pathways should be flushed 
with extract solution. 
Note: Extract volume may need to be adjusted depending on 
medical device size.
v. Place all pieces into the 20 x 150 mm tube containing 15 mL 
of preheated (37°C ±1°C) 0.01% SLS rinse solution. 
vi. Vortex the tubes for 30 – 60 seconds or until all pieces of the 
device are immersed in the rinse solution. 
vii. Sonicate the test containers for 60 minutes (wattage range 
150 – 480 watts) at 37°C ±1°C. Do not sonicate more tubes 
than can be vortexed within 15 minutes of completion of the 
sonication. Make sure the water in the sonicator covers the 